Welcome to the Validated Biosystems Website Newsletter, dedicated to serving the downstream processing community.
This CatoSource Newstand version will give you an idea of what Validated Biosystems has for downstream professionals.
Regular Features:
The Consultant: Experienced advice on key processing issues.
Solid Gold: User reviews of exceptional products and services.
The Jungle:
Practical tips on a range of downstream processing techniques
and resources.
S.W.A.T.:
Discussion of real-life
reader-contributed downstream processing problems, and suggestions for
their resolution.
Suppliers' Top Picks: Suppliers highlight their best process resources.
Free Offers: Free biotech promotional items on the web.
The Consultant
Every issue we'll address a practical topic in downstream processing, on
which we'll share insights and technical tips we've developed from over
15 years of hands-on idustrial process development. There will be a strong
emphasis on the steps required to achieve the best results, along with whatever
theory is necessary for it all to make sense. We welcome follow-up discussion,
which we will post in the next issue. Our topic for this issue is:
Linear and Step Gradient Elution; Data versus Dogma
Pete Gagnon, VBI
The relative value of linear and step gradients remains a point of controversy
in purification process design. Like most such controversies, its persistence
reflects incomplete articulation of the merits and limitations of the two
formats. In this article we'll discuss some of the major process parameters
affecting or affected by gradient format, with the goal of revealing how
the 2 formats can be applied most productively.
Product concentration
Step gradients have a reputation for eluting product at higher concentrations
than linear gradients. They do so in many contexts, but not all, and there
are limits to their concentrating ability in any case. Figures 1 and 2 contrast
peak volume as a function of step or linear gradient interval. As shown,
gradient format is a minor determinant of peak volume. The key factor is
the magnitude of the step or slope. Depending on sample composition and
resolution requirements, peak volume from linear gradients can be competitive
with peak volume from steps.
As implied by the above Figures, there is an inverse relationship between
peak concentration and resolution (Figures 3,4). The sacrifice of resolution
to achieve a high product concentration is generally more severe with step
gradients. Within a linear gradient, the relative relationships among the
eluting proteins tend to be well-preserved. In a step gradient, setting
a broader interval to achieve higher product concentration automatically
compromises purity. Such intervals are feasible with step gradients only
when the requirement for resolution is low.
It is important to look beyond the method at hand when evaluating resolution
requirements. If the contaminants flanking the product in one method are
easily removed by another method in the same process, then resolution requirements
for method at hand are low, despite the flanking contaminants (Figure 5).
Broad gradient steps can be employed to elute the product at high concentration
without imparing overall process performance.
Figure 6 illustrates the opposite situation where flanking contaminants
are shaired by a pair of separation methods. Linear gradients would be challenged
by such a situation, but step gradients would be wholly unsuitable. This
highlights the point that the foundation of a good process is built on complementarity
of separation methods. High resolution linear gradients can be used to maximize
the degree of complementarity, but they are not a substitute for the lack
of it.
When developing gradient specifications for step gradients, set the broadest
interval that doesn't compromise overall process performance. With linear
gradients, set the steepest slope. These actions will yield the highest
eluting product concentration for whichever format you use.
Eluted product concentration is limited by a number of other factors, regardless
of gradient setpoints. One of the most important is diffusional limitations.
The slow diffusion constants of proteins makes peak volume a function of
media particle and pore size distribution. No matter how extreme your elution
step, a given gel will always have a fixed minimum peak width, as a function
of particle and pore size distribution. Packing quality will have an effect
and dependency on diffusion makes peak width strongly dependent on flow
rate as well. Figure 7 illustrates increasing peak width in a linear gradient
as a function of flow rate for BSA on a HIC matrix and an anion exchanger.
The differences in the relative response, despite column dimensions, sample
load, particle and pore size distribution all being identical, demonstrate
the influence of other factors. In this case, the higher viscosity of the
high-salt HIC buffer was judged to be the dominant cause. However, differential
kinetics of the respective elution mechanisms cannot be discounted.
Product purity and recovery
Product purity within a given method is seldom as good with step gradients
as with linear gradients, and when it is, it's usually is achieved at the
expense of recovery. Narrowing the gradient intervals to partition out flanking
contaminants almost always requires sacrificing the leading or trailing
fractions of the product peak (Figure 8). This re-emphasizes the importance
of maximizing complementarity among process methods as the foundation to
process development.
Where high resolution is required, linear gradients are the best option.
Figure 9 illustrates a frequent pattern in linear gradient development.
As gradient slope is reduced, initially, resolution increases more than
peak volume. This reflects the rate of change in mobile phase composition
coming into phase with the kinetic limitations of the ligand:protein interaction.
With further slope reduction, peak volume increases more than resolution.
This doesn't mean that resolution won't continue to improve, just that you
will pay an increasingly high price for it. The transition point in resolving
efficiency versus peak volume can be estimated by comparing the ratio of
product peak height to adajcent valley height among chromatograms with different
gradient slopes.
Process reproducibility
Linear gradients have the ability to buffer minor process variations. So
long as the product elutes near gradient center, and the gradient amplitude
exceeds the range of process variation, external variations cause little
more than a modest deviation of gradient slope. The relative relationships
among the eluting proteins remain relatively unchanged. If uncontrolled
external process variation is high, maintaining the slope while extending
the gradient start and endpoints increases its insulating capability. Even
substantial variances are absorbed with little consequence. This is important
because the sources of variation are diverse and many of them are substantial.
One such source is variation in the fluidics architecture of process chromatographs,
especially as they compare with process development systems. Systems vary
with respect to accuracy of both flow and solvent proportioning. Equally
important, they vary with respect to the amount of internal solvent mixing
that occurs between the proportioning valve and the column. Each systems
has a characteristic "dispersion volume" -- the volume of solvent
required for complete transition from one gradient setpoint to another.
The larger the dispersion volume, the larger the volume of solvent required
to achieve a programmed setpoint. The effects on step gradients can be devastating.
Figure 10 illustrates process variation resulting from differences in dispersion
volume relative to column volume. The process was developed with a small
column on a chromatograph with a high dispersion volume. During development,
the wash step never reached target concentration within the programmed volume.
When the process was scaled to a larger column on the same system, the dispersion
to column volume ratio diminished, the wash step did reach its programmed
value, and the product eluted prematurely.
Degree of column loading also has disproportionate importance for step gradients.
Figure 11 illustrates variation in peak width and elution position as a
function of column load. Not only does the peak become wider with increasing
load, it elutes earlier. Step specifications set at a given column load
are valid only at that load. This is a particular problem in situations
where the product concentration and its proportion to contaminants in the
feedstream vary from lot to lot.
This is also an impediment to process development. Development columns must
be loaded to their intended process capacity throughout process development.
This is a circular trap since capacity varies according to the run conditions.
Setpoints for linear gradients, on the other hand, can be set preliminarily
with low subcapacity column loads, then adjusted to compensate for the load-shift
after other process specifications have been set. This is much simpler and
it conserves sample.
Other external variations also have significant impact on the efficacy of
step gradient setpoints. Hydrophobic interaction and protein A separations
are very sensitive to temperature. Variations of a few degrees can render
steps invalid, sacrificing purity, recovery, or both. Ion exchange is sensitive
to minor variations in conductivity. As with column load, these effects
have process development as well as reproducibility ramifications. The process
must be modeled, and setpoints validated across the range of process variation
that may affect the process. If resolution requirements are very permissive
then broad steps pose no serious reproducibility concern. Otherwise, the
"buffering capacity" of linear gradients makes their use essential.
Process sequencing
Step gradients offer process sequencing opportunities that linear gradients
rarely match. For example, you can often elute product from a HIC column
with a low salt buffer, and proceed directly to an ion exchanger with little
intermediate sample re-equilibration. Products eluting within a linear gradient
are likely to have a higher salt content, requiring either a higher degree
of dilution or complete buffer exchange. The same principle applies to other
process sequences.
Process control
Step gradients on ion exchangers can cause gross pH aberrations within the
column. With anion exchangers, a large step in chloride concentration can
liberate a sufficient concentration of hydroxide to raise the local pH to
12 and potentially denature the product. Acidification by hydronium ion
displacement can occur on cation exchangers. This puts more constraints
on buffer formulation to ensure adequate pH control. The gradual increase
of salt in linear gradients avoids this problem.
Process monitoring
Steps provide no information as to the composition of a peak. Three gradient
steps taken on a column loaded with a complex mixture will produce 3 peaks,
regardless of the gradient intervals. This makes extensive secondary testing
essential during process development, and also means that large-scale process
failures can be masked. Even linear gradients can't indicate the complete
composition of a peak, but the relationship among eluting peaks does provide
an index that allows immediate visual assessement as to whether or not the
process is within specified control limits. Linear gradient profiles also
make it possible to abbreviate the requirement for secondary testing during
method devlopment.
Process simplicity
The purported simplicity of step gradient applies to mechanical simplicity
only. When large-scale chromatography systems were limited to simple switch
valves this was an overiding factor, but no longer. Virtually all of the
current generation large scale systems have linear gadient capability equivalent
to the most sophisticated HPLCs.
With the wide availability of large-scale linear gradient chromatography
systems, step gradients have become more -- not less -- complicated than
linear gradients. The complications begin in development, as noted above,
where setting reproducible specifications requires comprehensive full-load
scale modeling. Accommodating all of the factors requires tedious balancing
and rebalancing of the step intervals to support the best combination of
purity, recovery, product concentration, and reproducibility. With linear
gradients, once the slope is defined, accommodating external process variation
is a simple matter of extending the start and endpoints sufficiently to
insulate the "core" segment.
Process economy
The higher resolution supported by linear gradients frequently allows purifications
to be conducted with fewer methods. A pair of linear methods will often
support purification performance equivalent or better than a triplet of
step methods, and triplets of linear gradient methods consistently outperform
quads of steps. This is an important distinction for process economics.
It reduces media requirements. It reduces column hardware requirements.
It reduces labor. It reduces storage space requirements. It means that expensive
manufacturing space is tied up for shorter periods per product -- thereby
increasing facility capacity -- and it reduces validation requirements.
The economic advantage is amplified by simpler development and better reproduciblity.
Conclusions
Gradient elution is the means by which the inherent complementarity among
separation methods is exploited to its greatest advantage. Step gradients
may be preferable where the relative selectivities among separation methods
make high resolution fractionation unnecessary. The more permissive the
fractionation requirements for a given method, the steeper the steps, the
higher the eluted product concentration, and the less the results will be
affected by external process variation. This tends to favor steps in processes
with more methods, and where external sources of process variability are
tightly controlled, as with manufacture of injectable products.
Linear gradients are generally a stronger option where resolution requirements
are high, where external process variables are poorly controlled, where
time pressure requires accelerating the development cycle, and where there
is economic pressure to minimize the number of fractionation methods. This
combination of requirements is more characteristic of in vitro reagent manufacturing
environments and preparation of investigational materials.
In practice, every purification represents a unique challenge, as well as
a unique set of opportunities. No preconceived philosophy about gradient
modes is going to give you the flexibility you need to achieve the best
process performance. Evaluate both formats, and apply them as they serve
you best.